Development and Implementation of a New Telerehabilitation System for Audio-visual Stimulation Training in Hemianopia
This study was conducted with an earlier version of AvDesk.
Summary
Article describing a new audio-visual telerehabilitation system for visual field defects and the first results in 3 adults who used it at home. All three detected more stimuli in the blind field when they could move their eyes, with effects still present 6–12 months later.
Objective
To describe the system and assess its feasibility and preliminary efficacy for home use under remote supervision.
Study design and protocol
- The system: roll-up semicircular panel with 24 LEDs in two rows and 12 loudspeakers, from 8° to 88° on each side (180° coverage); a camera that monitors head and gaze and pauses the stimuli if fixation is not maintained; wireless button; tablet for the patient; the therapist adjusts the programme remotely.
- Study type: preliminary study in 3 adult patients, IRCCS Fondazione Stella Maris (Pisa), with results analysed for each patient individually.
- Patients: a man who had a stroke at 71 (right inferior quadrantanopia), a young adult who underwent surgery for epilepsy (complete right hemianopia), a 40-year-old man who underwent surgery for meningioma (partial left hemianopia); all more than one year after the lesion.
- Training: at home, at least 5 days a week, in 6 stages in which the interval between sound and light decreases from 500 to 0 ms; all completed it within 5 weeks.
- Outcome measures: visual and audio-visual stimulus detection tests, with fixed eyes and with eye movements; computerised perimetry.
- Assessments: before, immediately after and at 6, 9 or 12 months depending on the patient.
Main results (as reported by the authors)
- Feasibility: all three set up and used the system at home on their own; the only technical issue reported was the button battery.
- Stimuli detected in the blind field with eye movements: significant increase in all three patients, still present at follow-up.
- With fixed eyes: significant improvement only in the first patient.
- Perimetry: unchanged. The training works on compensation through eye movements, not on the extent of the visual field.
Limitations
- Only 3 patients, analysed individually, without a control group.
- The assessment tests were performed with the same apparatus used for training.
- Preliminary study: the authors present it as a first step towards studies in larger samples.